BS-600M
Chemistry Analyzer
Technical Speci cations
System Function:
Throughput: 600 photometric tests per hour, up to 800
tests per hour with ISE
Measuring principles: Absorbance Photometry,
Turbidimetry
On-board tests: up to 77 photometric tests + 3 ISEs+ 3
serum indices
Methodology: End-point, Fixed-time, Kinetic, ISE
Sample Handling:
SDM: 120 samples by 12 racks, sample continuous
loading, STAT in priority
Sample volume: 1.5~45μL, step at 0.1μL
HbA1c sample: Whole blood for HbA1c test
Sample probe: Liquid level detection, clot detection and
collision protection
Reagent Handling:
Reagent tray: 80 positions with cooling system (2~8°C)
Reagent volume: 10~200μL, step at 0.5μL
Reagent probe: Liquid level detection, bubble detection
and collision protection
Built-in Bar Code Reader:
Sample and reagent bar code readers support Codabar,
ITF (Interleaved Two of five), Code 128, Code 39,
UPC/EAN and Code 93;
Capable to connect with LIS in the bi-directional mode
Reaction System:
Reaction volume: 70~300μL
Reaction temperature: 37°C with 0.1°C fluctuation
Reaction cuvettes: 124 reusable cuvettes with 8-step auto wash
Optical System:
Light source: Tungsten-halogen lamp
Photometer: Grating photometer
Wavelength: 16 wavelengths (340nm, 380nm, 412nm,
450nm, 480nm, 505nm, 546nm, 570nm, 605nm, 630nm,
660nm, 700nm, 740nm, 770nm, 800nm, 850nm)
Absorbance range: 0~3.5Abs(10mm conversion)
Control and Calibration:
Calibration mode: Linear(one-point, two point and
multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponen
tial, Polynomial, Parabola
Control rules: Westgard multi-rule, Twin plot
Operation Unit:
Operation system: Touch screen(optional), Windows
10(64bit)
Interface: RS-232, Network Port, USB/ parallel port
Working Conditions:
Power Supply: 200~240V, 50/60Hz, 1700VA
or 110/115V, 60Hz, 1700VA
Temperature: 15~30°C
Humidity: 35~85%
Dimension: 1380mm(length)*860mm(width)*1200mm(high)
Weight: 350kg
Flexible Sample Management
- Sample batch transfer
- Continuous sample loading by racks
- STAT Priority
- Quick rerun
HbA1c Tests with Whole Blood
- NGSP certi ed and IFCC traceable
- Onboard hemolysis without sample preparation
- No signi cant interference from Hb variants
- Lower detection limit of Hb, less rerun rate
Integrated ISE Module
- 1 blood tube for electrolytes and photometric chemistry tests
- long-life of ISE electrodes
- Easy operation and maintenance
Reliable Testing System
- World-class and excellent sigma metric performance with original reagent
- Intelligent abnormal reaction detection, alarming and auto-rerun to minimizes erroneous results
Flexible Sample Management
Continuous Sample Loading
Sample continuous loading by racks
Up to 120 sample positions
long walk-away time
Quick Start for Sample Rerun
Auto start for sample test
Automatic priority for STAT and rerun
STAT Priority
Dedicated STAT position, 1 minute aspiration
Quick report ISE+CO2 results in 7 minutes
Smart Dilution
Satented smart dilution function
Faster report
HbA1c Tests with Whole Blood
Onboard hemolysis with whole blood to improve working e ciency
No manual pretreatment to reduce biohazardous risks and human errors
120 T/H for HbA1c test to shorten turn-around-time
Mindray HbA1c reagent with enzymatic method can specifically detect glycated N-terminal amino acids from the hemoglobin beta-chains, which can effectively avoid interference from Hb variants (HbS, HbC, HbE and HbD) and other Hb derivatives (labile HbA1c, carbamylated Hb, and acetylated Hb).
Mindray enzymatic HbA1c gives reliable and cost efficient result with the first measurement and reduce further confirmations.
Lower Hb Detection Limit for HbA1c Test
Evaluate the Hb concentration in the process of hemolysis and start sample increasement tests in advance for low Hb concentration samples. This function can lower the detection limit and expand linearity range Hb to reduce the rerun rate and reagent waste for anemia samples.
Integrated ISE Module
Integrated ISE Module
- Report Na+, K+, CL
- 1 blood tube for electrolytes and photometric chemistry tests
Effcient and Easy operation
- Up to 300 tests per hour
- Up to 9 months of Na/K/Ref electrodes service time
- Automatic calibration
- Low consumption
- Easy operation and maintenance
Reliable Testing System
High Level Sigma Metric Performance
Original reagents and validated application
protocol lower the inaccuracy and impreci
sion which contributes to achieving
world-class and excellent sigma metric
performance.
Higher sigma level system can help laborato
ries offer high level of quality at reduced cost
and enhance clinician confidence and
satisfaction.
Intelligent Abnormal Reaction Detection
Powerful software with in-built algorithms that can monitor the reaction for abnormality, provide flags, and trigger auto-reruns to specimens with abnormal situations in test, such as substrate depletion, oxygen shortage, and hook effects, etc. This intelligent approach with detection, alarming and auto-rerun functions minimizes false results and the risk of a wrong diagnosis, and make reported test result accurate and reliable.